User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP E1356A:Strain Gage Multiplexer DATE: 29-Sep-94 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 95 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP Turn off HP 75000 mainframe. 1.004 DISP Remove HP E1356A 8-CH 350ê STRAIN RELAY MUX 1.004 DISP card from the rear of the mainframe. 1.004 DISP 1.004 DISP Remove the plastic protective cover from the 1.004 DISP mux card exposing the teriminal blocks. 1.005 HEAD {} 1.006 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.007 HEAD {º Bridge Completion Resisitor Test º} 1.008 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.009 HEAD ÍÍ Bridge Completion Resisitor Test ÍÍ 1.010 JMP 2.001 1.011 EVAL 2.001 DISP Set the Datron 1281 up to be able to measure 2.001 DISP the resistance at -E1 and -E2 on the mux card, 2.001 DISP in a four wire configuration. 2.001 DISP Connect the Datron to CH0 -E1 and -E2. 2.002 IEEE [@1281]OHMS AUTO,FWR 2.003 IEEE [@1281][D2000]RDG?[I] 2.004 ACC 2000 350.000Z 6P% 0.3P/ 2.005 MEME 2.006 MEMC Z 0.1000% chÿ0 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 3.001 DISP Connect the Datron to CH1 -E1 and -E2. 3.002 IEEE [@1281]OHMS AUTO,FWR 3.003 IEEE [@1281][D2000]RDG?[I] 3.004 ACC 2000 350.000Z 6P% 0.3P/ 3.005 MEME 3.006 MEMC Z 0.1000% chÿ1 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 4.001 DISP Connect the Datron to CH2 -E1 and -E2. 4.002 IEEE [@1281]OHMS AUTO,FWR 4.003 IEEE [@1281][D2000]RDG?[I] 4.004 ACC 2000 350.000Z 6P% 0.3P/ 4.005 MEME 4.006 MEMC Z 0.1000% chÿ2 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 5.001 DISP Connect the Datron to CH3 -E1 and -E2. 5.002 IEEE [@1281]OHMS AUTO,FWR 5.003 IEEE [@1281][D2000]RDG?[I] 5.004 ACC 2000 350.000Z 6P% 0.3P/ 5.005 MEME 5.006 MEMC Z 0.1000% chÿ3 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 6.001 DISP Connect the Datron to CH4 -E1 and -E2. 6.002 IEEE [@1281]OHMS AUTO,FWR 6.003 IEEE [@1281][D2000]RDG?[I] 6.004 ACC 2000 350.000Z 6P% 0.3P/ 6.005 MEME 6.006 MEMC Z 0.1000% chÿ4 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 7.001 DISP Connect the Datron to CH5 -E1 and -E2. 7.002 IEEE [@1281]OHMS AUTO,FWR 7.003 IEEE [@1281][D2000]RDG?[I] 7.004 ACC 2000 350.000Z 6P% 0.3P/ 7.005 MEME 7.006 MEMC Z 0.1000% chÿ5 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 8.001 DISP Connect the Datron to CH6 -E1 and -E2. 8.002 IEEE [@1281]OHMS AUTO,FWR 8.003 IEEE [@1281][D2000]RDG?[I] 8.004 ACC 2000 350.000Z 6P% 0.3P/ 8.005 MEME 8.006 MEMC Z 0.1000% chÿ6 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 9.001 DISP Connect the Datron to CH7 -E1 and -E2. 9.002 IEEE [@1281]OHMS AUTO,FWR 9.003 IEEE [@1281][D2000]RDG?[I] 9.004 ACC 2000 350.000Z 6P% 0.3P/ 9.005 MEME 9.006 MEMC Z 0.1000% chÿ7 #! Test Tol 0.35, Sys Tol 0.0027, TUR 129.630 (>= 4.00). 10.001 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 10.001 DISP {º THIS COMPLETES THE VERIFICATION º} 10.001 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 10.002 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.